User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: American-Reliance PPS-1302A:DC Power Supply DATE: 31-May-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 48 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R N P F W 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 DISP The following test equipment is required: 1.004 DISP 1. Metered vaiac DTB # 2-172 1.004 DISP 2. Transistor Devices DLP 50-150-3000A 1.004 DISP 3. Sensitive Research 20 Amp Shunt DTB # 14-109 1.005 DISP Set the UUT front panel controls as follows: 1.005 DISP POWER ON/OFF ....................................... OFF 1.006 DISP Connect the System port 1 to the UUT HP-IB connector. 1.007 DISP All connections are made to the rear panel 1.007 DISP terminal strip. 1.008 DISP ÿ Connect the UUT and the 1281 as follows: 1.008 DISP ÿ[27][91]1m 1302A TO 1281[27][91]m 1.008 DISP ÿ OUTPUT (+S) ÄÄÄÄÄÄÄÄÄÄÄÄ> INPUT-HIGH 1.008 DISP ÿ OUTPUT (-S) ÄÄÄÄÄÄÄÄÄÄÄÄ> INPUT-LO 1.009 DISP Connect UUT AC line cord to Metered Variac. 1.010 DISP Set the variac to 115 VAC. 1.011 DISP Set the UUT front panel controls as follows: 1.011 DISP POWER ON/OFF ........................................ ON 1.012 IEEE [D3000] 1.013 CALL American-Reliance PPS-1302A: Sub1 1.014 CALL American-Reliance PPS-1302A: Sub2 1.015 IEEE [SDC] 1.016 HEAD 1.017 DISP Set the UUT front panel controls as follows: 1.017 DISP POWER ON/OFF ....................................... OFF 1.018 HEAD {} 1.019 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.019 DISP {º THIS COMPLETES THE VERIFICATION º} 1.019 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.020 END #! T.U.R.s less than 4.00: 1 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 10 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.